Introduction
Turbidity Sensor adopts 90°scattering light principle, RS485 and standard Modbus, with a good advantage of wide range, self-cleaning and free maintenance; It can be used to measure turbidity value of backwash water in filter tank, raw water and sedimentation tank in waterworks, monitoring the water quality in industrial production and regular water monitoring for surface water, groundwater and pipe network.
Characteristics
Self-cleaning window film coated to reduce microbial contamination;
Self-cleaning wiper can secondary clean window surface;
Emits 880nm near-infrared light, can compensate for the influence of sample color;
Continuous measurement, fast response, suitable for continuous monitoring;
Many ranges for choice, can meet different accuracy requirements;
Flexible installation and easy for operation
SL61-A Low Range Turbidity Sensor
Specification | Details |
Size | φ48mm*195mm (length) |
Weight | 0.32KG |
Main Material | 316 Stainless Steel + POM |
O ring: viton ring+ rubber ring |
Cable: PVC |
Waterproof Rate | IP68 |
Measuring Range | 0~10NTU |
Temperature: 0~50℃ |
Measuring Accuracy | ±0.1NTU |
| Temperature: ±0.5℃ |
Resolution | 0.01NTU |
Temperature: 0.1℃ |
Data Output | RS 485, Modbus |
Power | DC 12~24V, average current < 30mA |
Storage Temperature | -20℃~+60℃ |
Working Temperature | 0~50℃ |
Pressure Range | ≤1Mpa |
Calibration Way | One point or two point calibration |
Cable Length | 5m standard cable, other length can be optional; |
Guarantee Period | 12 months |